APD-100 Automatic Powder Dispenser User Instructions 27
5.1 Setting Parameters
7. Type in a value between 1 and 100 and press the ENTER key to accept the
value displayed…
BEF. XX is displayed. Use this parameter to set the percent of the
weight at which the motor starts dispensing in pulse mode. XX is the
value of the setting. Settings are in percentages of the total powder
weight wanted for the charge so values can be between 1 and 100. The
default setting is 95.
This means the motor will run at pulse speed from the 95% weight level
to the 100% weight level.
Figure 5.3 Motor speed chart
8. Type in a value between 1 and 100 and press the ENTER key to accept the
value displayed…
ERRX.XX is displayed. This is an error parameter. Set this to an amount
between 0.3 grains (default) and 3.00 grains and when the scale is
finished auto-dispensing, if the weight on the scale varies from the
desired amount by more than this weight an ERROR message will
appear on the screen.
The equation below shows the relationship between the three parameters. CHG must
be less than or equal to SWI which must be less than or equal to BEF which must be
less than or equal to 100%.
CHG. XX <
SWI. XX < BEF. XX < 100%.
Also refer to the motor speed chart in Figure 5.3.